Electrospinning

TL Electro-spinning & Spray Equipment TL-PRO-BM

Product No :

TL Electro-spinning & Spray Equipment TL-PRO-BM

REQUEST A QUOTE

 

List of TL-PRO-BM:

  • 1. High voltage power supply device:
    0~50KV, digital display, over-current & over-voltage protection. Output power<25W.
    -20KV~0, digital display, over-current & over-voltage protection. Output power<25W.
    Option: -50KV~0, for conjugated electrospinning.
    2. Four-channel pump: can be controlled independently for individual syringe.
    Special designed for electrospinning. Easy to use. Digital display. Flow rate, volume and time of infusion can be controlled precisely. Compatible to work under high voltage.
    3. Drum collector(1-5000rpm, +/- 1rpm), 1pc; disk collector, 1pc. Plate collector with holder,1pc. Grooved collector, 1pc. 6 Nos of mandrel collector.

    4. Spinnerets with various holders.
    4.1 Single needle with holder
    4.2 Coaxial spinneret: Coaxial spinneret can produce Hollow nanofibers/microfibers, Core/Sheath nanofibers/microfibers. Using the coaxial electro-spinning method, different characteristics from each polymer can be combined into one fiber.
    4.3 Side by side nozzle
    4.3 Triaxial spinneret.
    4.4 Tri-channel spinneret.
    4.5 Multi-nozzle spinnerets:
    5. Light source: Special designed light source, LED, cold light source, very useful for observing the spinning process by naked eyes. 6. 6. Exhaust fan.
    This device is equipped to cool down the equipment in working process, and help the solvent vaporization. Press the ventilation button on the control panel to start working. This device is equipped to cool down the equipment in working process, and help the solvent vaporization. Press the ventilation button on the control panel to start working.

    7. Safety Measures: Emergency Stop Button; Door safety lock. (The high voltage power will be shut down automatically when door opened. ); Static removing system.

    8. Heater
    The syringe pump should be put outside when heater is using, otherwise, the syringe pump will be damaged. The best working temp. is 45~50 degree C. Please increase the temp. slowly when heating. In winter, heater is quite useful for solvent volatilization.

    9. Slide unit for spinnerets
    10. Tubeless spinneret
    11. Syringe & soft tube
    Plastic syringe: 100 Nos
    Needle: 100 Nos
    PTFE tubing with PP lure lock connector: 30 Nos
    PTFE tubing with PEEK lure lock connector: 45m

    12. Insulation X-Y platform for flat collector.

    Designed for electrospinning process, Compatible to work under high voltage. The flat collector placed on the platform can move under software control. This X-Y platform is quite useful for Near-field Electrospinning, NFES.

    13. Auto-clean device for needles

    14. Other accessories.
    15. Humidity control 

    As the solvent vapor should be removed from the chamber continuously, the inside humidity of the chamber can’t be controlled directly. While the following methods can be used to adjust the humidity of the room, as the air in the room will be withdrawn into the chamber continuously, the inside humidity of the chamber can be controlled accordingly. 
    1. Using an Industrial dehumidifier to reduce the level of humidity in the room.  
    2. Using an Industrial humidifier to increase the level of humidity in the room. 
    3. Insert dry air to adjust the humidity in the chamber. 
    Besides, when it is wet day, the heater in the chamber can be used with exhaust fan together to adjust the humidity within some extend. 

    15. Note: Grounding equipment should be prepared by the user. Specifications and design are subject to change without notice.